SUD-30X Cold Dry Block Temperature Calibrator
Business Type | Manufacturer, Exporter, Supplier, Trader |
Model | SUD-30X |
Control Range | -30 to 100 °C @ 25°C |
Resolution | 0.1 °C |
Control Accuracy | £0.1°C |
Thermal Stability | +0.01°C at -30°C, +0.02°C at 50°C, +0.04°C at 100°C |
Thermal Uniformity (Radial) | +0.04°C at -30°C, ,+0.08°C at 50°C+0. 10°C at 100°C |
Hysteresis | 0.05 °C |
Control Accuracy
£0.1°C
Thermal Stability
+0.01°C at -30°C, +0.02°C at 50°C, +0.04°C at 100°C
Thermal Uniformity (Radial)
+0.04°C at -30°C, ,+0.08°C at 50°C+0. 10°C at 100°C
Hysteresis
0.05 °C
Stabilization Time/settling time
10 minutes after set point is achieved.
Cooling Time
30 Min ( Ambient to -30°C) @25°C
Heating Time
15 Min (Ambient to 100 °C ) @25°C
Power Supply
230 V(AC), (50 Hz)
Power Consumption
300 Watt
Display
LCD, °C or °F user-selectable
Construction Size (H x W x D)
340(H) x 165(W) x 290(D) mm
Instrument Weight
Kg
Environmental operating conditions
0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
Immersion (Well) Depth
120 mm
Insert Length
120 mm
Insert OD
22mm
Method of Control
Self tunned PID controller
Specifications valid in environmental conditions
5°C … 25°C